Shared Flashcard Set

Details

Calibration of the Audiometer
Instrumentation and Electroacoustics
63
Audiology
Graduate
02/20/2017

Additional Audiology Flashcards

 


 

Cards

Term
Purpose of Calibration of the Audiometer:
-Verify that test equipment and signals are still
performing within ___ tolerances
-Verify ___ among test signals that produce comparable test results
-Correct for ___ that have occurred since the last check
Definition
manufacturer / standard; consistency; changes
Term
____ ____-____ presents all aspects of the standards for audiometers including distortion, allowable frequencies, values for zero HL for various transducers, etc.Developed in 1969, revised in 1989, 1996, 2004, and 2010
Definition
ANSI S3.6 – 2010: American National Standard Specification for Audiometers
Term
ANSI S3.6-2010:
-Consistent with ___ and ___ standard procedures. Ensures that all audiometers meet the same specifications ___ in the world they are manufactured and/or used
-Describes processes for ___ ear (rather than real
ear) evaluation of audiometer function
Definition
IEC; ISO; anywhere; mechanical
Term
Measures the electrical potential difference between two points in an electric circuit
Definition
voltmeter/multimeter
Term
What are the 3 different domains of calibration checks?
Definition
intensity, frequency, and time
Term
The voltmeter/multimeter...
-Can measure ___ output from an electrical device
-Can be used to determine the location of electrical ___
Definition
voltage; problems
Term
Device for measurement of sound pressure level. Used to measure SPL in a coupler or in sound field
Definition
sound level meter
Term
ANSI S1.4-1983 (R2006) classifies SLMs on the basis of features:
-Type 0 – ___ ___, laboratory use
-Type I – ___ ___
-Type II – ___ ___
-Other types –___
Definition
high precision; very precise; general purpose; hobby
Term
For audiometer calibration, at minimum a Type __ with one-third octave band filter and ___ microphone should be used
Definition
1; condenser
Term
Chambers used to simulate the acoustical and physical properties of a human ear
Definition
acoustic couplers
Term
Acoustic Couplers:
-Used to connect headphone/earphone to ___
instrument (typically, a sound level meter)
-Standardized ___ and ___-Enclose __ cubic cm of air in the ear canal when you put a supra-aural headphone on
-May contain a calibrated ___
Definition
measurement; shape; volume; 6; microphone
Term
simulates human ear up to 10,000 Hz. 3 cavities with 3 different volumes. Could be used for circumaural headphones
Definition
IEC 318
[image]
Term
encloses 2 CC of air approximating what would be between an insert and the TM. Used for inserts
Definition
Hearing aid couplers (HA-1 and HA-2)
[image]
Term
Coupler Applications-Artificial Ear:
-Enables measurement of ___ from air-conduction transducers (headphones and insert earphones)
-Exerts appropriate ___ to hold the headphone in place
-Interchangeable couplers allow measurement from ___ transducers
-Rubber feet minimize ___
Definition
output; force; multiple; vibration
[image]
Term
-The goal of the output check is to make sure the attenuator dial setting matches actual ___ level of test signals as closely as possible
-Output intensity checks are conducted for ___ transducers
-Output levels are expected to ___ from transducer to transducer; therefore, there is no ___ meaningful output level from the audiometer to measure
-Transducer-___ check
Definition
intensity; all; vary; single; dependent
Term
What does RETSPL stand for?
Definition
Reference Equivalent (Threshold) Sound Pressure Level
Term
-___ is the sound pressure level measured at a reference point which corresponds to (is equivalent to) threshold.
-Ex: If threshold (zero SL) is 16.5 dB SPL for Transducer X at Y Hz, then the RETSPL for that transducer, at that frequency, in that test condition, is ___
Definition
RETSPL; 16.5
Term
-Equipment: ___ microphone, ___
-Place earphone on ___ with a weight on top of it (for supra-aural headphones)
-For inserts, receiver tube is connected to adapter ___ of HA-2 coupler (no weight needed)
-Output is read as __, converted to dB SPL (re: 20 μPa), then compared to expected values at each frequency
-Reference SPL values provided in ISO 389-1 and ANSI S3.6
Definition
condenser; couplers; coupler; nipple; V
Term
When output has been established:
1. Attempt to bring audiometer into calibration using ___ ___
2. Use ___ sheet if necessary
3. If audiometer is off by more than __ dB at any frequency or by __ dB at three or more frequencies, contact service technician for recalibration and transducer check
Definition
trim potentiometers ; correction; 15; 10
Term
-Goal of Linearity Check: ___ signals should be attenuated by the dial in as linear a fashion as possible (i.e., turning the dial X degrees always changes the attenuation by Y dB)
Definition
Output
Term
Linearity Check:
-Transducer-independent check – linearity may be evaluated ___ (directly from the audiometer) or ___
(through a transducer)
-___ measurement: supra-aural earphone placed on an 6-cc coupler connected to a SLM with a one-third octave band filter
-___ measurement: voltmeter connected between audiometer output and earphone using a patch cord; electronic load in line to simulate earphone impedance
Definition
electronically; acoustically; Acoustical; Electrical
Term
Linearity Check
-Turn audiometer to maximum output, then attenuate in 5-dB steps until output can no longer be ___
-Acoustical technique: unit measure is dB ___
-Electrical technique: unit measure is ___
Definition
measured; SPL; voltage
Term
Linearity Check:
-Tolerance: decreases should be linear within __% of the interval step or __ dB, whichever is smaller
-Example: for a 5 dB step size setting, each attenuation should be between __ and __ dB
Definition
30; 1; 4; 6
Term
Frequency Calibration Equipment:
The ___ allows visual inspection of
signal voltages, usually as a graph of one or more electrical potential differences as a function of time
Definition
Oscilloscope
Term
The oscilloscope allows user to....
-Determine the ___ of an input signal
-Determine the ___ ___ of an input signal
-Detect ___ in a signal
-Determine whether a signal is ___ or ___
-Measure ___/___ time of a transient signal
-Test for ___ in multichannel instruments
Definition
voltage; time value (duration); distortion; AC; DC; rise/fall; crossover
Term
Frequency Calibration Equipment:
-The ___ ___ is a device that stores and displays the number of times (cycles) a particular input has been received
-The ___ ___ is a transform
function (Fourier transform) to analyze the frequency composition of an input signal
Definition
frequency counter; spectrum analyzer
Term
Frequency Calibration Equipment:
-The ___ ___ is a device that measures an input voltage signal
and calculates the energy at a target frequency
-Remaining total energy in the input is considered ___ energy
Definition
distortion meter; distortion
Term
Goal of frequency check: Frequency readout should ___ actual frequency of test signals as closely as possible
Definition
match
Term
Frequency Check:
-Equipment: electronic ___ or ___ ___
-Electronic counter gives a precise reading of ___ frequency
-Oscilloscope can be used for this purpose but with less ___. Typically accurate only within ___%; 1-3% is desirable
Definition
counter; frequency analyzer; output; precision; 5
Term
Frequency Check:
-Connect electronic counter between audiometer ___ and ___ using patch cords
-Transducer-___ check
-Set attenuator dial so that output is sufficient to ___
the counter
-Measure frequency for each setting on the frequency ___
Definition
output; earphone; independent; activate; control
Term
Frequency Check-Standard tolerances for audiometer types:
-Type 1: +/- __% of nominal
-Type 2: +/- __%
-Type 3-5: +/- __%
Definition
1; 2; 3
Term
Goal of distortion check: Output signals should contain as little ___ energy (energy at frequencies other than the frequency of the test signal) as possible
Definition
distortion
Term
Distortion Check:
-Harmonic distortion should be checked through ___
-Equipment: SLM w/ ___ band filter, ___
-Distortion will appear as attenuation which is not ___, particularly at high output levels (___ dB and above)
Definition
transducers; octave; couplers linear; 90
Term
Distortion Check:
-Place earphone on appropriate ___ with microphone connected to a SLM with ___ octave band filter
-Adjust ___ dial to appropriate setting for frequency tested. Goal is to be out of the noise floor.
-With octave band filter adjusted to match frequency being tested, measure output ___. Change filter to ___ harmonic, measure, output. Repeat until all ___ frequency harmonics are measured.
-Tolerance: total harmonic distortion less than or equal to ___%
Definition
coupler; 1/3; attenuator; SPL; second; higher; 2.5
Term
FM Signal Check:
-FM capability is a requirement of Type __ and __ audiometers
-FM defined by three characteristics: ___, ___ ___, and ___ ___
-Transducer-___ check
Definition
1 and 2; carrier, modulation rate, and frequency deviation; independent
Term
-___ is the center or nominal frequency– the setting of the frequency dial
-Tolerance: +/- ___%
Definition
Carrier; 3
Term
___ ___ is the number of times the FM signal completes a modulation cycle per second
-Tolerance: +/- ___%
Definition
Modulation rate; 10
Term
The ___ ___ frequency range over which change occurs for each modulation
-Tolerance: +/- ___%
Definition
Frequency deviation; 10
Term
Time Calibration:
-___ ___ ___ is a device for recording signal level (in dB) over time
-Provides a visual representation of the ___ ___ in an electrical (or acoustic) signal
Definition
graphic level recorder; time variation
Term
___-___ ___ check is a check to determine the precision of the audiometer in rising to desired output levels and descending back to zero
Definition
Rise-Fall Time
Term
Rise-Fall Time Check:
-Goal I: Output signals should increase from zero to desired output level (rise time) at a ___ rate every time they are presented, and should decrease from test level to zero (___ ___) at a consistent rate
-Goal II: Rise and fall time should be ___, ___, and ideally ___ one another
Definition
consistent; fall time; short; reliable; match
Term
Rise-Fall Time Check:
-Equipment: ___
-____ routed directly from audiometer into oscilloscope
-Transducer-___ check
-Typically tested using a ___ Hz tone, but it is good practice to check all frequencies from 250 to 6000 Hz
Definition
oscilloscope; Output; independent; 1000
Term
Rise-Fall Time Check:
-The ___ ___ is ength of time in msec (after interrupter switch is activated) between when signal is at target intensity -20 dB and when signal is at target intensity -1 dB. Time it takes for signal to go from x-20 to x-1.
Definition
rise time
Term
Inverse of rise time-time from x-1 to x-20
Definition
fall time
Term
ANSI S3.6 – 2004 specifies rise/fall times should be between __ and ___ msec
Definition
20 and 50
Term
Pulsed Signal Check:
-Type 1 and 2 audiometers are required to have capability to produce ___ tones
Definition
pulsed
Term
Pulsed Signal Check:
-Rise and fall times should be between ___-___ msec
-Plateau of the pulse tone no less than ___ msec
-Intensity difference between “on” and “off” phases of each pulse must be at least ___ dB
-Transducer-___ check
Definition
20-50; 150; 20; independent
Term
What coupler is used for the calibration of bone oscillators?
Definition
Mechanical Coupler a.k.a. “artificial mastoid”
Term
When was the Mechanical Coupler invented?
Definition
1960
Term
Artificial Mastoid:
-Standardized mechanical impedance - inertial ___ simulates the mechanics of the human head
-Contains built-in ___ force transducer
-___ ___ holds the bone oscillator in place at a standard force which can be adjusted (typically) between 2 and 8 N
-Transduces the force from a bone-conduction oscillator to an ___ signal
Definition
mass; electromechanical; moveable arm; electric
Term
What is a popular type of artificial mastoid?
Definition
B&K 4930
Term
Bone oscillator calibration:
-Oscillator mechanism removed from the headband and placed on the ___ ___, using the weighted arm to create the appropriate pressure
-___ Newtons, +/- .5 for mastoid placement
-Forehead placement requires ___ force on the oscillator to obtain comparable thresholds
-A ___ ___ ___ is used to measure the output of the bone
oscillator and compared to normative values based on bone oscillator RETSPLs
Definition
artificial mastoid; 5.4; more; sound level meter
Term
Intensity and Frequency Checks:
-Similar method to calibration of ___-___ transducers
-Artificial mastoid output to the SLM with ___-___ filter
-all checks of bone oscillator are transducer-___
Definition
air-conduction; octave-band; dependent
Term
-Distortion in the bone oscillator may be measured by connecting the output of the artificial mastoid to a ___ ___
-Only need to check frequencies that are used for ___ ___ using the bone oscillator
-Tolerance: THD less than or equal to ___%
Definition
frequency analyzer; audiometric
testing; 5.5
Term
Real-ear calibration of bone conductors is not recommended because methods require assumption that AC and BC
thresholds are ___ for normative group and it does not account for ___ noise in test room
Definition
equivalent; ambient
Term
Masking:
-May be generated by an ___ device (masking generator) or, more often, by the ___
-Masking outputs through all transducers should be ___ after verifying that the transducers themselves are working properly
-It is necessary to ___ all masking noises used for audiometric testing. White noise, weighted random noise for masking of speech, and narrow band noise are all defined by ANSI ___-___
-In general, the same checks are used for masking noises as are used for other ___ signal outputs. Linearity check, output check, etc
Definition
external; audiometer; checked; calibrate; S3.6-2004; audiometer
Term
Narrow Band Noise Calibration:
-Bandwidths are measured using a ___ analyzer
-Calibrated to the ___ ___ for each frequency
-Critical bandwidth: NBN will just mask a tone with the same ___ as the NBN center frequency at an SNR of -4 dB
Definition
spectrum; critical bandwidth; frequency
Term
Narrow Band Noise Calibration:
-___ level (measured using an SLM with octave band filter) calibrated in dB EML
-___ ___ ___: SPL of a masking signal that masks a pure tone at 0 dB HL 50% of the time when the masking signal and pure tone are presented to the same ear
-ANSI S3.6 specifies that NBN masking must range from __ dB EML to maximum ___-___ HL for each frequency, but should not exceed 115 dB SPL at any frequency
Definition
Output; EML (Effective Masking Level); 0; air-conduction
Term
Narrow Band Noise Calibration:
-Linearity of NBN masking should be measured the same way ___-___ linearity is measured
Definition
pure-tone
Term
White Noise:
-Typically calibrated in dB ___, but can be calibrated in ___
-Frequency response of masker: white noise SPL should be within +/- __ dB of the level at 1000 Hz from 250 to 5000 Hz
-In other words, if the white noise is 45 dB SPL when measured
at 1000 Hz, it should be between ___ and ___ dB SPL at all other
frequencies between 250 and 5000 Hz
-Can be ascertained using an ___ ___ ___ + ___ ___ ___ or with a spectrum analyzer
Definition
SPL; EML; 5; 40 and 50; SLM; octave band filter
Term
White Noise:
-Output response of masker: measured white noise output should be within ___ to ___ of nominal output (number on the masking level dial)
Definition
-3 to +5
Term
-Transducer output levels should be checked every ___ months or more often if you have reason to believe the output has changed
-Old equipment should be checked at least ___ per year
Definition
three; twice
Term
Examples of red flags in audiometers:
-The same air-bone gap is seen in two ___ patients
-BC thresholds are significantly ___ than AC thresholds
-Audiometric thresholds do not seem to ___ with other test results or your subjective impressions
-Patient reports hearing ___ sounds (particularly a patient who has been tested in the clinic before)
Definition
successive; worse; agree; unusual
Supporting users have an ad free experience!