Term
The diameter plays a major role in determining the beamwidth. True or False |
|
Definition
|
|
Term
The thickness plays a role in determining the operating frequency for piezoelectric transducer. T or F? |
|
Definition
|
|
Term
Beamwidth is also called: |
|
Definition
Lateral Resolution, Side-by-side dimension, angular, azimuthal, transverse. |
|
|
Term
The depth of natural focus is affected by which 2 principles? |
|
Definition
- operating frequency - diameter of the crystal |
|
|
Term
what is the distance between the surface of transducer and the natural focus called? |
|
Definition
"focal depth" or near zone length "NZL" |
|
|
Term
What is Near Zone Length Equation? |
|
Definition
|
|
Term
Small diameter crystals are good for |
|
Definition
|
|
Term
Large diameter crystals are good for |
|
Definition
|
|
Term
What is relation of NZL and operation frequency |
|
Definition
|
|
Term
A natural focus occurs when the operating frequency is ____ or/and when the crystal diameter is ____ |
|
Definition
|
|
Term
where does the highest intensity occur? |
|
Definition
at the focus (shallower than the focus in reality) |
|
|
Term
|
Definition
thin layer, with impedance lower than that of piezoelectric crystal, but higher than that of tissue |
|
|
Term
an ideal diameter of matching layer |
|
Definition
|
|
Term
limitations of simple crystal |
|
Definition
high acoustic impedance, fixed focal depth, long SPL, no ability to steer |
|
|
Term
what does matching layer do? |
|
Definition
reduces acoustic mismatch between crystal (high) and tissue (low), thus improving transmission efficiency |
|
|
Term
|
Definition
ability to separate 2 structures in the axial (longitudinal, radial, depth, range) dimension |
|
|
Term
|
Definition
ability to separate 2 structures in the lateral (side by side, angular, transverse, azimuthal) dimension |
|
|
Term
axial resolution is determined by |
|
Definition
|
|
Term
lateral resolution is determined by |
|
Definition
|
|
Term
why increasing the frequency to achieve better AR is often not a solution? |
|
Definition
because of the required imaging depth (the size of patient), higher frequency leads to lower penetration |
|
|
Term
What is the best possible way to obtain the best AR in diagnostics? |
|
Definition
using backing material (shortening SPL by lowering number of cycles) |
|
|
Term
5 ways of changing the focus |
|
Definition
- mirror (not popularly employed) - lenses - curved elements - electronic focusing (requires multiple beams) - retrospective gating (requires multiple beams) |
|
|
Term
multiple transducer elements are called |
|
Definition
|
|
Term
disadvantages of lens in focusing |
|
Definition
- acoustic impedance mismatch - potential heating surface of transducer |
|
|
Term
curved surfaces, a newly employed technique of focusing, requires the use of |
|
Definition
composite material other than crystal |
|
|
Term
what are the 2 most useful techniques among the 5 techniques of focusing |
|
Definition
- electronic focusing - retrospective gating these are variable-focusing which allows user control |
|
|
Term
elevation resolution is determined by the |
|
Definition
beamwidth in the elevation plane |
|
|
Term
|
Definition
convert voltage (electropotential) into sound waves and vice versa |
|
|
Term
|
Definition
shortens SPL, improving longitudinal resolution |
|
|
Term
backing material is also called |
|
Definition
|
|
Term
|
Definition
minimize acoustic impedance mismatch between crystal and tissues |
|
|
Term
|
Definition
varying voltage which creates the transmit and is processed to create image with receive mode |
|
|
Term
|
Definition
transmit- delivers voltage to signal receive- delivers echo voltage to system receiver |
|
|
Term
|
Definition
helps to focus beam (2D arrays in elevation plane) |
|
|
Term
|
Definition
- Doppler only transducer, Blind, Pencil Beam transducer |
|
|
Term
|
Definition
- inexpensive - good sensitivity of Doppler - small footprint, allowing insonation in smaller areas |
|
|
Term
|
Definition
- no ability to produce an image - fixed focus |
|
|
Term
With Hanafy lens, the crystal is ___ along crystal edges, and ____ in the middle |
|
Definition
|
|
Term
The consequence of varying thickness in Hanafy lens is |
|
Definition
there are different resonance across crystal surface: higher frequency in the middle, lower frequency in the outer, thicker aspect -> resulting in a broader field of view in the elevation plane |
|
|
Term
2D arrays can steer and focus in both ____ and _____ planes |
|
Definition
|
|
Term
Another name for near field |
|
Definition
|
|
Term
Another name for far field |
|
Definition
|
|
Term
|
Definition
the area between transducer to beam focus |
|
|
Term
|
Definition
|
|
Term
|
Definition
where the beam reaches minimum diameter |
|
|
Term
Focal region is the region over which the beam is most tightly focused and also called |
|
Definition
|
|
Term
|
Definition
ability to distinguish between 2 objects in any of the three dimensions: axial, lateral, elevation |
|
|
Term
operating frequency is also known as |
|
Definition
|
|
Term
|
Definition
the center frequency of the transmit bandwidth |
|
|
Term
f0 (operating frequency) in CW is determined by |
|
Definition
frequency of the drive (transmit) voltage |
|
|
Term
f0 (operating frequency) in PW is determined by |
|
Definition
crystal thickness and speed of sound in crystal material.[image] |
|
|
Term
Electrical capacitance is defined as |
|
Definition
stored electrical charge when a potential difference (voltage) is applied between electrical conductors |
|
|
Term
|
Definition
Capacitive Micromachined Ultrasound Transducer |
|
|
Term
CMUT technology is based on the concept of creating and detecting sound through _____ |
|
Definition
|
|
Term
|
Definition
|
|
Term
Acoustic window in CMUT is |
|
Definition
a material put over the cells as a barrier between patient and CMUT cells |
|
|
Term
CMUT doesn't require backing material but requires matching layer. T or F? |
|
Definition
False. CMUT requires backing material, but doesn't require matching layer as it has acoustic window |
|
|
Term
what are electrodes in CMUT used for? |
|
Definition
Electrodes are basic components of capacitor, which hold electrical charge to create electric field |
|
|
Term
|
Definition
used to increase capacitance and keep electrodes from shorting in collapse mode |
|
|
Term
when an electric voltage is applied to the faces of either type, the thickness of the element increases or decreases depending on |
|
Definition
|
|
Term
other names of transducers |
|
Definition
probe, scanhead, transducer assembly |
|
|