Term
Voltage
Definition
Force (potential) causing electron movement; Higher voltage indicates a stronger ―push/pull‖ of electrons; Unit of measure: Volts (V)
Term
Current
Definition
Flow of electrons within a completed circuit path; Unit of measure: Ampere (amp) or miliamperes (mA); 1 amp = 1000 mA; Electrical symbol: I
Term
Resistance
Definition
Opposition to current (also referred to as impedance); Unit of measure: Ohm ; Electrical symbol: R or  ; Typical range: 300 – 100 Ohms
Term
Energy
Definition
Fundamental unit of work; Product of voltage, curent, pulse duration (pulse width); Unit of measure: micro Joules (J) ; J = 1000 mJ = 1,000,000 J ; Electrical symbol: E
Term
4 NBG codes
Definition
–Chamber paced
–Chamber sensed
–Response to sensing
–Rate modulation
Term
Pulse Generator
Definition
Contains the power source (battery) and the electronic circuitry necessary for operation of the device
Term
Sensing
Definition
Ability of the device to sense and respond to the electrical activity of the heart
Term
Interference
Definition
Any unwanted signal or noise sensed by the device
•May originate from:
–Electromagnetic interference (EMI)
–Myopotentials
–Tall P or T waves.
Term
Factors affecting resistance
Definition
-Lead insulation break
-Lead fracture
-Fracture of the lead wire, increases resistance and decreases current flow
Term
Normal Pulse width value
Definition
Normally 0.5 msec in permanent pacers and 2 msec in temporary pacers.
Term
Battery size in pacemakers
Definition
Usually 5 volts in permanent pacemakers and 15 volts in temporary pacers.
Term
Minimal electrical energy required to consistently capture the heart outside the refractory period.
Definition
Stimulation Threshold
Term
How is the Stimulation Threshold measured
Definition
measured at a pulse width of 0.5 msec
(always in volts)
Term
The stimulation threshold is at its lowest when?
Definition
during implant
Term
Peak Stimulation Threshold occurs when?
Definition
2 to 6 weeks after implant
Term
Exit block
Definition
Occurs when the stimulation threshold exceeds the output of the pulse generator.
Term
What will epinephrine do to the stimulation threshold?
Definition
it will lower the threshold
Term
Amplitude
Definition
The total height (both positive and negative deflection) in millivolts of the intracardiac signal.
AKA:
•Intrinsic deflection (ID) - the peak-to-peak deflection of the intracardiac signal as sensed by the pulse generator.
•The greater the amplitude or ID, the more likely it will be sensed.
•Usually only measurement sufficiently required for determining sensitivity.
Term
Usually only measurement sufficiently required for determining sensitivity.
Definition
Amplitude or Intrinsic deflection
Term
Slew rate
Definition
is the rate of change in voltage in an intracardiac electrogram and is mathematically represented by dv/dt.
Term
Factors affecting sensing
Definition
–Lead maturation
–Lead dislodgement/penetration
–Ischemia
–Lead integrity/design
–Noise
–Drugs
Term
All Pulse Generators must have at least these two circuits
Definition
-Timing
-Output
Term
Delivers a pulse at a constant preset rate, regardless of the intrinsic activity
Definition
Asynchronous generator
-Also called a Fixed-rate generator
Term
Allows pacing and sensing circuit parameters to be changed non-invasively.
Definition
Programming
Term
Refractory periods (in pacemaker terminology)
Definition
refers to the brief period after a sensed or paced beat in which the sensing circuit does not sense.
Term
Ability to interrogate the pacemaker and receive a readout of the programmed parameters and the status of the pacemaker.
Definition
Telemetry
Term
Most commonly used dual chamber pacemaker.
Definition
DDD
Term
Transcutaneous pacing (TCP):
Definition
–Also referred to as external cardiac pacing, noninvasive pacing, external transthoracic pacing, etc.
–Heart is stimulated with externally applied cutaneous electrodes that deliver an electrical impulse across the intact chest wall to the myocardium

•TCP consists of a external device connected to two electrodes with a cable/wire:
–One electrode placed on the right pectoral area next to the sternum (anterior) and one on the left scapular area (posterior).
–Also right pectoral area and the lateral portion of the chest wall (V5 or V6 position) may be used if access to the back is not feasible.
Term
Indicated in the treatment of hemodynamically significant bradycardia not responsive to atropine therapy.
Definition
Transcutaneous pacing (TCP)
Term
Indications for temporary pacing:
Definition
–Hemodynamic compromising bradycardia.
–Bradycardia with escape rhythms unresponsive to drug therapy.
–Symptomatic 1st and 2nd degree type I Heart Block.
–2nd degree type II and 3rd degree heart block
–Overdrive pacing of refractory tachycardia.
–Newly acquired BBB in presence of acute MI.
–Prophylactic use during PTCA and right heart cath in patients with LBBB.
Term
How do you find the stimulation threshold with pacing?
Definition
Stimulation threshold is the last mA that captured before loss. use 2 to 4 times the stimulation threshold.
Term
Complications of temporary pacing
Definition
–Hemothorax, pneumothorax, and thrombophlebitis.
–Malignant ventricular arrhythmia's.
–Myocardial wall perforation and cardiac tamponade
Term
ECG Guidance for placement of Temporary Pacemeakers -P wave
Definition
•Large and inverted in SVC
•Biphasic in mid RA
•Small in RV
•Negative in PA
Term
ECG Guidance for placement of Temporary Pacemeakers -QRS segment
Definition
•Small in RA
•Large in RV (ectopy common)
•ST segment: Will elevate when catheter is in contact with wall
Term
Advantages of transvenous pacemaker
Definition
–Ease of insertion.
–Decreased morbidity/mortality.
–Decreased surgical trauma.
–Reduced recovery time.
Term
Disadvantages of transvenous Pacemaker
Definition
–Exit block due to acute injury process
–Lead dislodgement.
Term
What is the most common approach in the packing community?
Definition
Endocardial Approach (Transvenous)
Term
Most common vessel for transvenous pacing lead?
Definition
–Cephalic (most common)
–External jugular
–Subclavian
–Internal jugular
–Femoral (least common)
Term
When is acute stimulation threshold measured?
Definition
Performed at implant
Term
When is chronic stimulation threshold measured?
slide 96
Definition
Performed 6 to 8 weeks post implant
Term
After finding your stimulation threshold, what should you then set the voltage output to?
Definition
The voltage output is then usually set 2 to 4 times higher than the measured threshold at implant to reduce the possibility of the pacer failing to capture due to fibrosis.
Term
For stimulation threshold Increased surface area equals:
Definition
–Increased sensing
–Increased threshold
–Less resistance (for the current to flow).
Term
Current sensitivity standard for most implantable pulse generators is what?
Definition
1.5 mV
Term
The peak to peak voltage of the ID, or amplitude, usually ranges from
Definition
5 - 15 mV (R wave).
Term
When unwanted signals such as T waves, P waves (unless pacing the atrium), EMI, pacemaker afterpotentials, myopotentials are sensed, the pulse generator is inappropriately inhibited.
Definition
Oversensing
Term
When the patient’s R wave as transmitted by the implanted lead does not display an amplitude that will inhibit the pulse generator.
Definition
Undersensing
Term
All functional parameters of the pulse generator should be analyzed prior to and after connecting to lead.
Definition
–Rate in pulse per minute (ppm) and msec intervals
–Pulse width in msec
–Pulse output in both V and mA under a 500ohm load
–Pulse generator sensitivity
–Refractory periods following both sensed and paced beats
–Hysteresis
Term
If a pulsed magnetic field is transmitted, it can influence a piece of metal called a ______________ to complete an electrical circuit.
Definition
reed switch
slide 122
Term
Defibrillation
Definition
Procedure by which direct current (DC) is delivered across the heart (transmyocardial current) for the purpose of converting unstable ventricular tachycardia or ventricular fibrillation to a sinus rhythm
Term
Most frequent rhythm seen in sudden cardiac death is what?
Definition
Ventricular Fibrillation
Term
Is a procedure by which direct current (DC) that has been synchronized to discharge with the QRS complex is delivered across the heart for the purpose of converting an supraventricular or unstable ventricular tachyarrhythmia to a sinus rhythm
Definition
Synchronized Cardioversion
Term
For synchronized cardioversion sedation is required unless what?
Definition
hemodynamically unstable
Term
Strength of defibrillation is most commonely expressed in?
Definition
energy (joules or watts-second) which is the product of power and duration
Term
For pediatrics... how many joules per point / per kg
Definition
1 joule per pound
2 joules per kg
Term
Firm pressure of the electrodes will decrease transthoracic resistance by as much as what for defib?
Definition
25%
Term
How much pressure due you apply to the electrodes for defib?
Definition
11kg (25lb) for each electrode
Term
You should attach an External Defibrillator (AED) only to patients that weigh more than what?
Definition
50kg or 110lbs because most AEDs have a minimum energy level of 200 Joules
Term
A device like a pacemaker which is simply an electrical circuit requiring a closed loop. The electricity is directed across a large portion of the myocardium in an effort to depolarize most of the ventricle.
Definition
Automated Implantable Cardiac Defibrillator
Term
Indications for Implantable Cardiac Defibrillator
Definition
–Intended for the use of patients whom are at high risk for sudden cardiac death due to ventricular arrhythmias
–Incessant VT or VF that is not suppressed by medical, surgical, or other ablative therapy.
–Expanded for asymptomatic non-sustained VT with decreased ventricular function
Term
Contraindications to an ICD
Definition
•Transient ventricular tachyarrhythmias due to reversible causes:
−Drug toxicity
–Electrolyte imbalance
–Sepsis
–Hypoxia
•Transient ventricular tachyarrhythmias due to other factors:
–Myocardial infarction
–Electrocution
Term
Defibrillation threshold (DFT)
Definition
is the lowest energy that successfully terminates ventricular fibrillation.
Term
What is the goal of DFT testing
Definition
The goal of DFT testing is to ensure that the maximal energy output of the device has an extremely high (> 99%) probability of terminating ventricular fibrillation in any given patient.
Term
Output
Definition
How much power is delivered
Term
Timing
Definition
Precisely when the pacemaker impulse will be delivered to the myocardium
Term
Also called a Fixed-rate generator
Definition
Asynchronous generator
Term
What does Sensing do?
Definition
–Monitors heart activity
–Acts as a receiver tuned only to wanted signals
•P waves—atria
•R waves—ventricles
–Other signals are discriminated against.
–Determines how the PG responds to a sensed depolarization
–May reset timing circuit during intrinsic heart activity
–May trigger the output circuit to pulse in synchrony with heart activity
Term
What are the pacemaker programmed parameters that you can check with telemetry?
Definition
–Rate
–Pulse width
–Output
–Sensitivity
–Refractory periods
Term
Transcutaneous pacing pulse width is factory set. What is the range?
Definition
20 to 40 msec
Term
What is the typical programmable range of beats per minute in a permanent pacemaker?
Definition
30 to 120 bpm
•Most pacemakers are preset by the manufacturer at 60 to 70 bpm.
Term
When in the respiratory cycle is transthoracic resistance at its minimal?
Definition
Full Expiration
Term
How does the duration of time in V-Fib affect electrical defibrillation?
Definition
The longer VF persists the less likely successful electrical defibrillation.