Term
|
Definition
responsible for producing x-rays and heat |
|
|
Term
|
Definition
accuracy with which a test reveals the patient's medical condition |
|
|
Term
|
Definition
how much clincally useful information the test provided |
|
|
Term
what are the components of an x-ray/radiographic system? |
|
Definition
1. x-ray tube & support system 2. collimator assembly 3. x-ray table 4. x-ray generator 5. control console 6. image receptor holder device |
|
|
Term
the image receptor may be a conventional film-screen cassette or more commonly a _______________ _____________ |
|
Definition
|
|
Term
what is the x-ray tube made of? |
|
Definition
|
|
Term
what are the primary components of a x-ray tube? |
|
Definition
|
|
Term
|
Definition
positive electrode of the x-ray tube |
|
|
Term
|
Definition
negative electrode of the x-ray tube |
|
|
Term
what supports the x-ray tube and allows you to position the x-ray tube over and around the patient? |
|
Definition
|
|
Term
what type of energy is supplied to the x-ray tube through the two large cables? |
|
Definition
|
|
Term
the x-ray tube converts electrical energy into |
|
Definition
|
|
Term
how can you control the number and energy of the x-rays produced by the x-ray tube? |
|
Definition
by adjusting the control console |
|
|
Term
what is directly below the x-ray tube? |
|
Definition
|
|
Term
what is the x-ray beam limiting device in the x-ray/radiographic system? |
|
Definition
|
|
Term
what is responsible for the x-ray field size? |
|
Definition
|
|
Term
|
Definition
increasing collimation=decreasing x-ray field size |
|
|
Term
where is a high intensity light bulb found? |
|
Definition
|
|
Term
when a light is projected on the patient what does this indicate? |
|
Definition
|
|
Term
|
Definition
positive beam limitation: automatically collimates x-ray field size to minimize patient exposure |
|
|
Term
why is it wrong to have an image receptor that is 10 x 12 and a collimation of 12 x 12? |
|
Definition
you are over-exposing the patient |
|
|
Term
what is the only reason to expose a patient to radiation? |
|
Definition
to obtain a useable image |
|
|
Term
what the most obvious part of a x-ray/radiographic system? |
|
Definition
|
|
Term
what are the 3 characteristics of an x-ray table? |
|
Definition
1. tilt a. tilting b. nontilting 2. height a. adjustable b. nonadjustable 3. movement a. stationary b. free-floating |
|
|
Term
explain a: a. 90-90 table b. 90-30 table c. 90-45 table |
|
Definition
a. 90-90 table: vertical to horizontal in either direction b. 90-30: 90 in one direction and 30 in another c. 90-45: 90 in one direction and 45 in another |
|
|
Term
|
Definition
easily transmits x-ray energy with very little absorption |
|
|
Term
TRUE or FALSE? x-ray tables are NOT radioluscent |
|
Definition
|
|
Term
|
Definition
located below the table top and holds the x-ray cassette (image receptor) stationary during x-ray exposure |
|
|
Term
|
Definition
|
|
Term
the Bucky mechanism tries to keep the image receptor centered on the __________ ___________ |
|
Definition
|
|
Term
what is the workhorse of the total system? |
|
Definition
|
|
Term
what are the 2 main components of the x-ray generator? |
|
Definition
electrical cabinet and operator console |
|
|
Term
the power of an x-ray system is expressed in |
|
Definition
|
|
Term
the control console allow you to |
|
Definition
1. turn the x-ray system on and off 2. select x-ray exposure factors 3. start and stop exposure 4. provides audible and visual signal that x-ray exposure is occuring |
|
|
Term
what are the 5 generic controls of the console? |
|
Definition
1. main power-on and off 2. kilovolt peak kVp 3. mA- milliamperage 4. timer 5. rotor-exposure switch |
|
|
Term
|
Definition
|
|
Term
|
Definition
|
|
Term
|
Definition
quality; kilovoltage peak; penetrating abilities of an x-rays |
|
|
Term
what is the range of kVp's in radiography? |
|
Definition
|
|
Term
|
Definition
quantity of x-rays produced |
|
|
Term
|
Definition
|
|
Term
|
Definition
|
|
Term
the device that begins the exposure is the |
|
Definition
|
|
Term
the rotor exposure has two switches |
|
Definition
1. prep switch 2. exposure switch |
|
|
Term
activation of the prep switch causes the |
|
Definition
anode to rotate and prepares the x-ray tube |
|
|
Term
how long does the prep phase take? |
|
Definition
|
|
Term
the termination of the exposure must be indicated both |
|
Definition
|
|
Term
the duration of x-ray exposures may be established _____________ or _______________ |
|
Definition
manually or automatically |
|
|
Term
|
Definition
automatic exposure control |
|
|