produce long range of grays |
|
|
Term
|
Definition
study of film emulsions response to exposure |
|
|
Term
|
Definition
beyond Dmax; image reversal |
|
|
Term
reading a sensitomety curve: steep slope indicates - affect on exposure factor |
|
Definition
fast film response, with little latitude, and higher contrast - decrease exposure *curve is on the left |
|
|
Term
|
Definition
leeway or margin of error one has with a given group of exposure factors - determined by kV (decreasing makes more of an effect) |
|
|
Term
Slow II screen speed -affect and use |
|
Definition
-200 -give better detail -good for extremities -increase recorded detail - increase dose |
|
|
Term
|
Definition
400 - used for thinker, denser structures, children -lower exposure factors |
|
|
Term
II screens - contrast and density affect |
|
Definition
-directly related to density - no affect on contrast |
|
|
Term
Grid ratio -contrast and density |
|
Definition
-inversely related to density -directly related to contrast |
|
|
Term
minimum response/reaction time |
|
Definition
length of shortest exposure possible with a particular AEC |
|
|
Term
AEC
error: terminating exposure too soon
-cause |
|
Definition
-tissue of less density than the one of interest is center over sensor or photocell
-sensor is not completely covered
|
|
|
Term
AEC
exposure terminates to late causing overexposure
- cause |
|
Definition
tissue of greater density is centered over photocell/chamber |
|
|
Term
AEC
adjusting exposure factor |
|
Definition
mAs can be adjusted by using master density control :plus/minus density, body habitus control, or measuring
- if overexposed because of an exposure shorter then minimum reaction time use lower mA station |
|
|