Give the traditional and SI unit for the measurement of radiation that ionizes air
|
|
Definition
roentgen (R) and couloumb/ kg |
|
|
Term
Give the traditional and SI unit for absorbed dose |
|
Definition
|
|
Term
Give the traditional and SI units for the measurement that take into account the Weighting factor for diferent types of tissues. This uses LET plus a wieghting factor |
|
Definition
|
|
Term
How is the tolerance in Kvp checked and what must this tolerance be within |
|
Definition
Wisconsin test casstte or digital, must be within 4 kvp |
|
|
Term
What is the tolerance for collimation |
|
Definition
must be within 2% of the SID |
|
|
Term
What is the tolernace of the effective focal spot size and how is it tested |
|
Definition
Slit camera, pinhole camera, star test pattern, 50percent |
|
|
Term
How is the accuracy of the timer tested fpr a single phase generator
|
|
Definition
spinning top - the arch will fraction of time |
|
|
Term
How is the accuracy of the three phase generator tested |
|
Definition
|
|
Term
This program involves all aspects of the radiology process |
|
Definition
|
|
Term
What process involves and ensures only the safe and reliable operation of equipment |
|
Definition
|
|
Term
What test ivolves using various mA, and time combos for a given mAs and may be used for testing adjacent mA stations. |
|
Definition
|
|
Term
What must exposure linearity tolerance be within |
|
Definition
|
|
Term
What does Exposure linearity use to test its tolerance |
|
Definition
|
|
Term
What equipment will be used for a Exposure reproducibility |
|
Definition
|
|
Term
What test will utilize kvp , mA and time controls for a given mAs and what is the tolerance |
|
Definition
Exposure Reproducibility , 5% |
|
|
Term
What tool is used to test AEC and verify reproducibility of exp |
|
Definition
|
|
Term
How is Flouro exp rate tested |
|
Definition
|
|
Term
What is the limit for entrance skin exposure |
|
Definition
|
|
Term
What is is the image intensifier or automatic brightness control cheked for |
|
Definition
to ensure rad dose to input phosphor is constant |
|
|
Term
What tool is used to check resolutin of a television system |
|
Definition
|
|
Term
The Actual Patient Dose as measured by a meter embedded in the collimator is called |
|
Definition
|
|
Term
QA Dictates Cassettes should be cleaned and inspected how often ? |
|
Definition
|
|
Term
QA Dictates that Cassettes should be erased how often |
|
Definition
|
|
Term
Digital Imaging is Driven by |
|
Definition
|
|
Term
What is the range of angles the grid can use called |
|
Definition
|
|
Term
What is it called when brigthness is maintained over a wide range of exposures |
|
Definition
|
|
Term
What Law does this Equation Represent
mAs=mAs |
|
Definition
|
|
Term
What is the Formula for the inverse square law |
|
Definition
Iold/Inew=Dnew2/Dold2
I=intensity D= SID |
|
|
Term
What is the Density Maintenace formula and when is it used |
|
Definition
mAs old/ mAs new = D old2/ D new2
Used for changes in SID to maintain density |
|
|
Term
What three factors Affect Subject Contrast |
|
Definition
Atomic Number
Tissue Density
THissue Thickness |
|
|
Term
|
Definition
Height of the lead strips / distance of the lead strips =
H/D
4:1 -6:1 |
|
|
Term
What other terms are used for grid radius and what is its definition |
|
Definition
focal distance , focal range
Distance at which the focused grid may be used |
|
|