Term
Which medical gases support combustion |
|
Definition
"Helium, Oxygen, Air, Nitrous oxide, Nitric oxide, Heliox (He/O2), CO2, " |
|
|
Term
Which gases are nonflammable |
|
Definition
"nitrogen (N), Helium (He), Carbon dioxide, O2, Air, NO, Heliox (He/O2), CO2" |
|
|
Term
How is water content removed from medical air |
|
Definition
|
|
Term
What is DOTC and ICC folowed by on the shoulder of a compressed gas cylinder |
|
Definition
(Interstate commerce commission) followed by cylinder classification such as 3A or 3AA and normal filling pressure in pounds per square inch |
|
|
Term
What is below DOT and ICC on the shoulder of a compressed gas cylinder |
|
Definition
"letter size of the cylinder (E, G, etc)" |
|
|
Term
WHat is on the third line below DOT and ICC on the shoulder of a compressed gas cylinder |
|
Definition
mark of ownership followed by stamp or mark identifying inspecting authority |
|
|
Term
WHat is on the opposite side of a compressed gas cylinder |
|
Definition
an abbreviation indicating the method of cylinder manufacture and information regarding the original safety test and dates of all subsequent tests |
|
|
Term
"What does the notation ""EE"" followed by a # indicate on the shoulder of a compressed gas cylinder" |
|
Definition
the elastic expansion of the cylinder in cubic centimeters under test conditions |
|
|
Term
What does an asterik next to the test date on the shoulder of a compressed gas cylinder indicate |
|
Definition
DOT approval for 10-year testing |
|
|
Term
what does a plus sign next to the test date on the shoulder of a compressed gas cylinder indicate |
|
Definition
the cylinder is approved for filling to 10% above its service pressure |
|
|
Term
What is something about compressed gas cylinders regarding coding |
|
Definition
it's color coded and labeled |
|
|
Term
How is duration of flow calculated from an H cylinder of oxygen |
|
Definition
duration of flow (min) = 3.14 (factor for H cylinder) * pressure/flow. Note that the factors for: G cylinder is 2.41; E is 0.28 and D is 0.16 |
|
|
Term
|
Definition
standards for threaded high-pressure connections between large compressed gas cylinders and their attachments |
|
|
Term
What are the large compressed gas cylinders |
|
Definition
|
|
Term
|
Definition
Pin indexing applies to valve outlets of small cylinders up to and including E. |
|
|
Term
What type of connection is used for PISS |
|
Definition
"yoke type where the upper yoke fits over the lower valve stem. Two pins, projecting from the inner surface of the yoke connector, mate with two pinholes bored into the valve stem. Proper pin position aligns the small receiving nipple of the hyoke with the recessed cylinder valve outlet" |
|
|
Term
|
Definition
"ASS and PISS provide standards for high pressure connections between cylinders and equipment, the DISS was established to prevent accidental interchange of low pressure (less than 200 psig) medical gas connectors. " |
|
|
Term
Where are DISS connections typically found |
|
Definition
"at the outlets of pressure-reducing valves attached to cylinders, at the station outlets of central piping systems and at the inlets of blenders, flowmeters, ventilators, and other pneumatic equipment" |
|
|
Term
WHat does the DISS system consist of |
|
Definition
"externally threaded body and a mated nipple w/ a nut. as the two parts are joined, the shoulders of the nipple and the bores of the body mate and the union is held tgether by a hand-tightened hex nut. indexing is achieved by varying the dimensions of the borings and shoulders" |
|
|
Term
"An intubated mechanically ventilated patient is using an HME and starts to develop thick, tenacious secretions. what action should be taken?" |
|
Definition
|
|
Term
WHat happens when a bourdon gauge encounters back-pressure resistence |
|
Definition
(bourdon gauge is a flowmeter device that is used in combination with an adjustable pressure-reducing valve). The pressure difference across the orifice and actual output flow decrease. the bourdon gauge flow reading depends on upstream pressure which stays constant. the gauge reading would be falsely higher than the actual delivered flow because it measures upstream pressure. the gauge registers flow even when the outlet is completely blocked |
|
|
Term
Why is oxygen monitored closedly with COPD patients |
|
Definition
"COPD patients with chronic hypercapnia ventilate less. Patients hypoventilate when given oxygen because of suppression of the hypoxic drive. Normal response to PCO2 is blunted. The primary stimulus to breathe being lack of O2 as sensed by peripheral chemoreceptors. The increase in O2 levels supresses chemoreceptors, depresses ventilatory drive, and elveates PCO2. High blood oxygen levels may disrupt the normal V/Q balance and cuse an increase in dead space to Vt ratio and an increase in PaCo2" |
|
|
Term
What action would you take when removing a flowmeter from the wall outlet and there is a sudden massive gas leak |
|
Definition
"try to reinsert and pull back out again due to a valve not properly closing. If this doesn't work, leave in the wall until it can be fixed" |
|
|
Term
What are the advantages of a SVN vs MDI |
|
Definition
"less patient coordination required, high doses possible, no CFC release" |
|
|
Term
what are some disadvantages of MDI vs SVN |
|
Definition
"Pt coordination required, Pt activation required, high percentage of pharyneal deposition, risk of abuse, difficult to deliver high doses, not all meds are available, most units still use ozone-depleting CFCs." |
|
|
Term
"How is total flow from an air entrainment calculated. EX. 50% O2, 15L/min" |
|
Definition
liters air/liters )2 = (100-%02)/(%02-21) ex. ansser is 41L/min |
|
|
Term
What are the clincial signs/indications of hypoxemia (abnormal deficiency of oxygen in arterial blood) |
|
Definition
"PaO2<60 mmHg or Sao2 <90% on room air; PaOc of SaO2 below desirable range for clinical situation; also, tachypnea - tachycardia - cyanosis, distressed overall appearance." |
|
|
Term
What is retinopathy of preamturity |
|
Definition
abnormal eye condition occurs in some premature or low-birth weight infants who receive supplemental oxygen. An excessive blood oxygen level causes retinal vasoconstriction which leads to necrosis of blood vessels. |
|
|
Term
"with retinopathy of prematurity, what happens in response to necrosis of blood vessels" |
|
Definition
new vessels form and increase in number. Hemmorhage of these delicate new vessels causes scarring behind the retina. |
|
|
Term
"in retinopathy of prematurity, what does scarring behind the retina often lead to" |
|
Definition
detachment and blindness. |
|
|
Term
Who does ROP affect most often |
|
Definition
"neonates up to approx. one month of age, by which time the retinal arteries have sufficienty matured" |
|
|
Term
What is Bernoulli's principle |
|
Definition
fluid's lateral pressure varies inversely with velocity |
|
|
Term
|
Definition
total gas pressure = sum of partial pressures |
|
|
Term
|
Definition
rate of diffusion of a gas is inversely proportional to the square root of its gram molecular weight |
|
|
Term
|
Definition
solubility of a gas in liquid is proportion to its pressure if temp is constant and the gas does not react with the liquid |
|
|
Term
|
Definition
Hyperbaric oxygen therapy and is the therapeutic use of oxygen at pressures greater than 1 atm. Pressures during HBO therapy usually are expressed in multiples of atmopheric pressure absolute (ATA) |
|
|
Term
|
Definition
|
|
Term
What pressures are more HBO therapy conducted at |
|
Definition
|
|
Term
What are the physiologic effects of HBO therapy |
|
Definition
"bubble reduction (boyle's law), hyperoxygenation of blood and tissue (henry's law), vasoconstriction, enhanced host immune function, neovascularizatoiin" |
|
|
Term
What acute conditions benefit from HBO therapy |
|
Definition
"decompression sickness, air or gas embolism, carbon monoxide and cyanide poisoning, acute traumatic ischemi, crush injuries and limb suturing, clostridial gangrene, necrotizing soft tissue infection, ischemic skin graft or flap" |
|
|
Term
What are some chronic conditions that benefit from HBO therapy |
|
Definition
"diabetic wounds of lower extrmeties and other non-healing wounds, refractory osteomyelitis, actinomyosis (chronic systemic abscesses), Radiation necrosis (as an adjunct to conventional treatment)" |
|
|
Term
Nitric oxide is used primarily for what purpose? |
|
Definition
To treat neonates with hypoxic respiratory failure with associated pulmonary hypertension. |
|
|
Term
"what are some potential, but currently not used uses for inhaled NO" |
|
Definition
"ARDS, persistent pulmonary hypertension of the newborn, pulmonary hypertension follwing cardiac surgery, cardiac transplantation, acute pulmonary embolism, COPD, congental diaphragmatic hernia, sickle cell disease, testing pulmonary vascular responsiveness" |
|
|
Term
|
Definition
|
|
Term
what kind of method is used to create aerosol in a USN |
|
Definition
|
|
Term
USN - how does the crystal transducer create aerosol |
|
Definition
"converts radio waves into high-frequency mechanical vibrations(sound). these virbrations are transmitted t aliquid surface wher ethe intense mechanical energy creates a cavitation in the liquid, forming a standing wave or geyser which sheds aerosol droplets. vibrational energy is transmitted either indirectly through a water-filled couplant reservoir or directly to a solution chamber. Gas entering the chamber inlet picks up the aerosol particles and exits through the chamber outlet" |
|
|
Term
What do the properties of the ultrasonic signal in a USN signal determine |
|
Definition
the characteristics of the aerosol generated by these nebulizers. |
|
|
Term
what does the frequenchy of the USN signal determine |
|
Definition
determines aerosol particle size which is inversely proportional to signal frequency |
|
|
Term
WHat does signal amplitude of USN signal determine |
|
Definition
"ignal aplitude directly affects the amount of aerosol produced. the greater the amplitude the greater the volume of aerosol output. Unlike frequency, signal amplitude may be adjusted by the clinician" |
|
|
Term
What is the treatment for laryngitis or croup |
|
Definition
cool mist nebulizer - mist tent |
|
|
Term
what are the indications for the use of humidity therapy |
|
Definition
"primary: humidifying dry medical gases, overcoming the humidity deficit created when the upper airway is bypassed. Secondary: managing hypothermia, treating bronchospasm caused by cold air" |
|
|
Term
|
Definition
"special-purpose large-volume nebulizer. used especially for the administration of ribavirinto infants with respiratory synctial virus infection. it incorporates a drying chamber with its own flow control to produce a stable aerosol. it reduces medical gas source from teh normal 50 pounds per square inch gauge (psig) line pressure to as low as 26 psig with an adjustable regulator. used to treat infants with RSV, respiratory synctial virus treated ribaririn" |
|
|
Term
what does one cubic foot of liquid oxygen equal |
|
Definition
|
|
Term
WHat devices are capable of delivering 100% relative humidity at BTPS |
|
Definition
|
|
Term
Which oxygen delivery device most likely to be tolerated by children |
|
Definition
Oxygen enclosure such as oxygen tent |
|
|
Term
What is the main problem with oxygen tents |
|
Definition
large swings in oxygen concentration due to frequen opening and closing of the canopy |
|
|
Term
What is the best method of administration of controlled oxygen therapy to infants |
|
Definition
|
|
Term
How is oxygen delivered to an oxygen hood |
|
Definition
it is delivered to the hood through either a heated air entrainment nebulizer or a blending system iwth a heated humidifier. a minimum flow of 7 L/min should be set to prevent accumulation of CO2. |
|
|
Term
WHat flow would need to be used to maintain stable high oxygen concentration in an oxygen hood |
|
Definition
|
|
Term
WHat may cause stress on neonatal patients with regard to an oxygen hood |
|
Definition
flows greater than 10-15 L/min due to harmful noise levels |
|
|
Term
What must be precisely maintained in an oxygen hood for an infant |
|
Definition
the temperature of gases provided to an infant in an oxygen hood should be precisely set to maintain a neutral thermal environment (NTE). The NTE varies with weight and age. |
|
|
Term
|
Definition
plexiglas enclosures that combine servo-controlled convection heating with supplemental oxygen |
|
|
Term
WHat is the best way to control oxygen delivery to infants in an incubator |
|
Definition
|
|
Term
"With an infant in an incubator using an oxyhood, where must hte oxygen concentration and gas temperature be assessed" |
|
Definition
within the oxyhood and not the incubator |
|
|
Term
What devices would most likely cause over hydration when used on a continuous basis |
|
Definition
large volume jet nebulizers and USNs |
|
|
Term
|
Definition
liquid particles suspended in a gas |
|
|
Term
what is the most common device used to generate bland aerosols |
|
Definition
large-volume jet nebulizers |
|
|
Term
how are large-volume jet nebulizers powered |
|
Definition
pneumatically powered attaching directly to a flowmeter and compressed gas source |
|
|
Term
How are liquid particle aerosols generated in large-volume jet nebulizers |
|
Definition
"by passing a gas at a high velocity through a small jet orifice. The resulting low pressure at the jet draws fluid from the reservoir up to the top of a siphon tube, where it is sheared off and shattered into liquid particles. The large, unstable particles fall out of suspension or impact on the internal surfaces of the device, including the fluid surface (baffling)" |
|
|
Term
"With a large-volume jet nebulizer, how would heat be added" |
|
Definition
"as with humidifiers, if heat is required, a hot-plate, wrap-around, yolk collar, or immersion element can be added." |
|
|
Term
How are heated large-volume jet nebulizers different from humidifiers with regard to the control of delivery temperature |
|
Definition
"they rarely have servo-controlled systems to control delivery temperature. many systems do not even shut down when the reservoir empties, resulting in the delivery of hot, dry gas to the patient. failure of the element can also cause a loss of heating capacity without warning to the clinician" |
|
|
Term
How are pulse demand delivery devices advantageous |
|
Definition
it synchronizes O2 delivery with inspiration to conserve O2 (uses approx 60%) less |
|
|
Term
|
Definition
"(American National Standards Institute) The institute oversees the creation, promulgation and use of thousands of norms and guidelines that directly impact businesses in nearly every sector: from acoustical devices to construction equipment, from dairy and livestock production to energy distribution. They set the minimum level of humidty for intubated patients at 30 mg/L" |
|
|
Term
|
Definition
"specs on gas cylinders, regs for safety testing" |
|
|
Term
|
Definition
National fire protection agency. Compressed Gas Agency. They set cylinder storage guidelines |
|
|
Term
what are the guidelines that apply to cylinder storage |
|
Definition
"store in racks or chain to wall; other than the wooden racks, store no other combustible material; store away from sources of heat; store flammagle gases separately from those that support combustion; keep protective cap in place; don't store compressors and cylinders together; contain and store cylinder supply systems with one hour fire resistive rating; segregate full and empty cylinders; warning sign regarding oxidizing gas and alert against smoking; cool, well-ventilated area; " |
|
|
Term
what are the guidelines that apply to cylinder transport |
|
Definition
"use cylinder carts with securing mechanism; keep cylinder caps in place; protect cylinders from striking other cylinders or objects; avoid dropping, dragging, etc; make sure appropriately labeled" |
|
|
Term
what are the guidelines for cylinder use |
|
Definition
secure to pt bedside that prevents cyl from falling; do not use flammable materials; never cover with any material; open cyl valve slightly to remove dust before attaching to regulator; never use cyl that need repair; do not alter markings/color; never place near heat; never secure to movable objects; make sure connection is appropriate (ASSS for H and G and PISS for E); post no smoking sign |
|
|
Term
what are the variable FIO2 low-flow O2 delivery systems used in RT |
|
Definition
"nasal cannula, nasal catheter, transtrachial catheter" |
|
|
Term
What are the fixed FIO2 low-flow O delivery systems employed in RT |
|
Definition
"AEM, AE neb, blending system, isolette, incubator" |
|
|
Term
what are the controls of an ultrasonic nebulizer |
|
Definition
amplitude - increase to increase output and flow - increase to decrease density |
|
|
Term
What would be provided to a pt receiving nasal O2 at 3 L/min complaining of nasal dryness and irritation |
|
Definition
add a bubbler ( not generally required for flows less than 3 L/min) |
|
|
Term
What features of an SVN affect its performance |
|
Definition
"baffles, fill volume, residual drying volume, position, continuous vs intermittent, reservoirs/extensions, vent/valves/gas entrainment, tolerance in manufacturing" |
|
|
Term
How do you minimize a patient's infection risk between drug treatments wiht an SVN |
|
Definition
rinse with sterile H2O and air dry |
|
|
Term
particle distributions produced by a typical MDI |
|
Definition
|
|
Term
what time period shoudl elapse between each puff of a single MDI treatment |
|
Definition
|
|
Term
Primary safety concerns in the application of hyperbaric oxygenation |
|
Definition
"most frequent is barotrauma: ear or sinus trauma or tympanic membrane rupture; alveolar overdistension and pneumo, gas embolism. oxygen toxicity: CNS reaction, pulm reaction; fire, sudden decomp, reversible vision changes, claustrophobia, decreased cardiac output" |
|
|
Term
what factor is used to compute the duration of flow for an E cylinder |
|
Definition
|
|
Term
what factor is used to compute the duration of flow for an H cylinder |
|
Definition
|
|
Term
what factor is used to compute the duration of flow for a D cylinder |
|
Definition
|
|
Term
what does the plus sign on a cylinder mean |
|
Definition
approved for filling 10% over its service pressure |
|
|
Term
what is the humidity output of a simple unheated humidifier |
|
Definition
|
|
Term
"when monitoring a pt receiving O2 therapy, what needs to be assessed" |
|
Definition
"cardiac, pulm, and neuro stats" |
|
|
Term
what is the simplest way to increase the humidity output of a humidifier |
|
Definition
|
|
Term
double the length of an oxygen tube would have what effect on resistance to flow |
|
Definition
|
|
Term
what is the approximate half life of carboxyhemoglobin |
|
Definition
"room air: 5 hours, 100% fio2: 80 minutes; 3 ATA HBO: 23 min" |
|
|
Term
when is a continuous nebulizer appropriate treatment |
|
Definition
|
|
Term
what are the characteristics of servo-controlled heated humidifier |
|
Definition
"adjusts heater power automatically, through feedback from thermistor placed at or near pt airway. provides automatic heater shutdown, includes temp safety alarms." |
|
|
Term
what is the key functional difference between disposable partial rebreathing and nonrebreathing O2 masks |
|
Definition
nonrebreathers have one-way insp and exp valves |
|
|
Term
How would you clean a cylinder valve outlet of foreign material |
|
Definition
|
|
Term
what are the advantages of passover humidifiers |
|
Definition
"advantages over bubblers are: maintain sats at high flows, add little or no resistane to spntaneous breathing circuits, do not generate aerosols" |
|
|
Term
for what should sputum collected by aerosol therapy induction be inspected |
|
Definition
"for diagnosis of TB, pneumocystis carinii pneumonia, lung cancer" |
|
|
Term
compressed gas cylinders are manufactured from what mainly |
|
Definition
|
|