Shared Flashcard Set

Details

Ultrasound Physics
ch.8 Transducer, bandwidth, & Q-factor
39
Other
Undergraduate 1
02/03/2021

Additional Other Flashcards

 


 

Cards

Term
The fundamental or operating frequency is..
Definition
affected by the thickness of the PZT element (thickness=1/2 wavelength)
Term
How thick is the element if the frequency is 3MHz?
Definition
half the wavelength of the wave in the PZT
Term
The piezoelectric effect is best described as
Definition
deformation from voltage sent to the crystals
Term

The impedance of the PZT crystals is higher than the impedance of skin tissue

(TRUE or FALSE)

Definition
TRUE
Term

Which of the following could not be used as the coupling medium (gel)?

a. saline solution

b. water

c. barium sulfate

d. aqueous gels 

Definition
c. barium sulfate (doesn't allow propagation of US wave)
Term

The gel has greater impedance than the matching layer but less than the impedance of the PZT elements.

(TRUE or FALSE)

Definition
FALSE
Term
Highest to Lowest impedance.
Definition
PZT, Matching Layer, Gel, and Skin
Term

The damping layer decreases the PRP at a specific imaging depth.

(TRUE or FALSE)

Definition
FALSE
Term

The backing material decreases the Q factor.

(TRUE or FALSE)

Definition
TRUE
Term
If the transmitted pulse has frequencies from 3.5 MHz to 6.5 MHz, what is the bandwidth?
Definition
3.0 MHz
Term

The backing material decreases the SPL. 

(TRUE or FALSE)

Definition
TRUE
Term
Which is NOT a result of damping layer?
Definition
Sensitivity Increases
Term
Which affects the frequency of the transmitted beam from CW ultrasound?
Definition
Ultrasound System Electronics
Term

In PW ultrasound, the Q-factor should be made larger as possible.

(TRUE or FALSE)

Definition
FALSE
Term
Which is most important in producing the frequency of the beam in soft tissue.
Definition
Propagation speed of the PZT elements
Term

If the frequency of the electrical excitation voltage of a pulsed wave transducer is 6 MHz, the main or center frequency of the transducer is 6 MHz. 

(TRUE or FALSE)

Definition
FALSE
Term

The diameter of the active element of a transducer helps to determine the frequency of the US produced by the transducer.

(TRUE or FALSE)

Definition
FALSE
Term

Two piezoelectric crystals are made from the same material. The thicker crystal will make a pulsed transducer with higher frequency.

(TRUE or FALSE)

Definition
FALSE
Term

For a pulsed wave transducer, if the PZT crystal thickness is doubled, then the frequency of sound is halved.

(TRUE or FALSE)

Definition
TRUE
Term

The acoustic impedance of the matching layer equals the acoustic impedance of skin.

(TRUE or FALSE)

Definition
FALSE
Term

A broadband transducer is more characteristic than a narrow bandwidth for imaging transducers.

(TRUE or FALSE) 

Definition
TRUE
Term

A very high Q-factor transducer is more common than low Q-factor for diagnostic imaging.

(TRUE or FALSE)

Definition
FALSE
Term
With a pulsed wave transducer, what occurs if the PZT thickness is halved? (Decreased)
Definition
Frequency is doubled. (Increased)
Term
With a pulsed wave transducer, what occurs if the PZT thickness is doubled? (Increased)
Definition
Frequency is halved. (Decreased)
Term

The damping material in a transducer acts to: increase the pulse length.

(TRUE or FALSE)

Definition
False
Term

The damping material in a transducer acts to: improve the system's longitudinal resolution.

(TRUE or FALSE)

Definition
True
Term

The damping material in a transducer acts to: decrease the quality factor. 

(TRUE or FALSE)

Definition
True
Term
What is NOT a natural element?
Definition
Lead titanate zirconate. (man-made) manufactured
Term
Mechanical energy into electrical energy and vise versa is:
Definition
Piezoelectric property
Term

Transducers are cleaned with high temperature germicide. 

(TRUE or FALSE)

Definition
False
Term

Using a transducer with a cracked plastic covering may shock patient.

(TRUE or FALSE)

Definition
True
Term

Damping or backing material reduces the # of cycles per pulse and cuts down the ringing.

(TRUE or FALSE)

Definition
True
Term

Damping layer creates longer pulses with good quality factor.

(TRUE or FALSE)

Definition
False
Term

Damping material reduces sensitivity and efficiency of transducer

(TRUE or FALSE)

Definition
True
Term

The backing material decreases the duty cycle at a particular PRF.

(TRUE or FALSE)

Definition
True
Term

The pulse duration increases because of the backing material.

(TRUE or FALSE)

Definition
False
Term
What would cause the PZT properties to be eliminated?
Definition
High temperatures
Term
The measurement that relates bandwidth to the operating frequency is...
Definition
Q-factor
Term
The ____ of the transducer is the ratio of the operating frequency to the bandwidth.
Definition
Q-factor